• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL STEM FIBER METAL TAPER COLLARLESS 12/14 NECK TAPER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL STEM FIBER METAL TAPER COLLARLESS 12/14 NECK TAPER; PROSTHESIS, HIP Back to Search Results
Catalog Number 65786201320
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Unspecified Infection (1930); Insufficient Information (4580)
Event Date 09/19/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-02814, 0001822565-2023-02816, 0001822565-2023-02817.G2: australia.H10: cat #: 00801803201 / femoral head sterile product do not resterilize 12/14 taper / lot #: 60441090.Cat #: 65620005221 / shell porous without holes 52 mm o.D.With calcicoat ceramic coating / lot #: 60378385.Cat #: 00632005032 / liner 20 degree elevated rim 32 mm i.D.For use with 50/52/54 mm o.D.Shells / lot #: 60482805.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event Description
It was reported a patient underwent a hip revision for unknown reasons.Attempts have been made and no further information is available.
 
Event Description
It was reported that a patient underwent a hip revision approximately seventeen years post implantation due to an infection.All the components were removed and replaced.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM FIBER METAL TAPER COLLARLESS 12/14 NECK TAPER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17949375
MDR Text Key325838774
Report Number0001822565-2023-02815
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K980711
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Catalogue Number65786201320
Device Lot Number60287094
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/17/2023
Supplement Dates Manufacturer Received11/08/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age88 YR
Patient SexFemale
-
-