|
Model Number CDDRA500Q |
Device Problem
Product Quality Problem (1506)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient presented for routine implant of the implantable cardioverter defibrillator.During the procedure the intended implant device was observed to have a cloudy header.The device was not implanted, and the procedure was completed with a replacement.The patient was stable.
|
|
Manufacturer Narrative
|
Correction: the reported event of header discoloration was confirmed.However, the cloudy appearance was found to be acceptable since all the header components and the tip of the test leads were visible through the header.Interrogation of the device revealed the device was above the elective replacement indicator (eri) when received.The device was tested on the bench with leads and resistor loads.Telemetry, sensing, pacing, low voltage impedance, high voltage (hv) impedance, patient notification, and hv shock were tested.No anomaly was detected.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
|
|
Manufacturer Narrative
|
The reported event of header discoloration was confirmed.However, the cloudy appearance was found to be acceptable since all the header components and the tip of the test leads were visible through the header.Interrogation of the device revealed the device was above the elective replacement indicator (eri) when received.The device was tested on the bench with leads and resistor loads.Telemetry, sensing, pacing, low voltage impedance, high voltage (hv) impedance, patient notification, and hv shock were tested.No anomaly was detected.
|
|
Search Alerts/Recalls
|
|
|