ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ALLURE QUADRA RF CRT-P; NO MATCH Back to Search Results
Model Number PM3242
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2023
Event Type  Injury  
Event Description
It was reported that patient presented for routine generator change procedure.During procedure it was found that patient's atrial lead and right ventricular lead was not able to disengage from device header it was further noted that set screws of atrial and right ventricular port were stripped.The device was remained implanted.There were no patient consequences.
 
Event Description
New information received notes that the device was explanted and replaced.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17951315
MDR Text Key326352198
Report Number2017865-2023-49988
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507912
UDI-Public05414734507912
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2016
Device Model NumberPM3242
Device Lot NumberA000002527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET.; TENDRIL.; TENDRIL.
Patient Outcome(s) Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-