Model Number HL-90 |
Device Problems
Fluid/Blood Leak (1250); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that during testing the front cover was leaking, and the circulation pump was loud.No patient involvement was reported.
|
|
Manufacturer Narrative
|
Other, other text: b3: date of event and d4: udi number is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
Other, other text: d10, h3 and h6 - evaluation codes: updated device evaluation: one device was returned for investigation.Visual inspection noted the device received in poor condition.Front cover has multiple scratches.Microswitch is missing.Line cord discolored and worn.Quick connect was corroded.Water tank was cracked on the top and bottom.Out dated printed circuit board (pcb) and power switch.Enclosure was cracked under quick connect.Functional testing found water was leaking from the water tank cover and from the crack under the quick connect.The complaint was confirmed.Root cause was attributed to too much torque when tightening the water tank screws and the quick connect nut.What or who caused this condition was not established.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
|
|