Model Number HL-90
Device Problems Fluid/Blood Leak (1250); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during testing the front cover was leaking, and the circulation pump was loud.No patient involvement was reported.
 
Manufacturer Narrative
Other, other text: b3: date of event and d4: udi number is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: d10, h3 and h6 - evaluation codes: updated device evaluation: one device was returned for investigation.Visual inspection noted the device received in poor condition.Front cover has multiple scratches.Microswitch is missing.Line cord discolored and worn.Quick connect was corroded.Water tank was cracked on the top and bottom.Out dated printed circuit board (pcb) and power switch.Enclosure was cracked under quick connect.Functional testing found water was leaking from the water tank cover and from the crack under the quick connect.The complaint was confirmed.Root cause was attributed to too much torque when tightening the water tank screws and the quick connect nut.What or who caused this condition was not established.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17953927
MDR Text Key326041867
Report Number3012307300-2023-09711
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/17/2023
Supplement Dates Manufacturer Received11/03/2023
Supplement Dates FDA Received11/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-