• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 SYSTEM CONTROLLER; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 106531US
Device Problems Low Audible Alarm (1016); Alarm Not Visible (1022); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Event Description
It was reported that the controller was exchanged in clinic because the pump indicator light was gone and the speakers were muffled.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported events of dim pump running leds and the audio being muffled were not confirmed.Additional information provided communicated that no products would be returned for analysis.The root cause of the reported events could not be conclusively determined through this analysis.The device history records were reviewed and the records revealed that the heartmate 3 system controller, serial number (b)(6), was manufactured in accordance with manufacturing and qa specifications.The system controller was shipped to the customer on 09mar2020.Heartmate 3 instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate 3 patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment, including the controller.Heartmate 3 patient handbook section 6 ¿ "caring for the equipment" describes how to care for and clean all equipment, including the system controller.This section instructs the user to, at least monthly, inspect the system controller¿s audio sounders for dirt or grease.If a change in tone or in loudness is noticed during a system controller self-test, the audio speaker sockets may be obstructed.Audio speaker sockets may be cleaned using a small cotton swab that is moistened (not dripping) with rubbing alcohol.Never insert anything sharp (like a toothpick or pin) into the sounder holes.This can damage the speakers inside.The heartmate 3 patient handbook and the heartmate 3 instructions for use caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 SYSTEM CONTROLLER
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key17954373
MDR Text Key326385246
Report Number2916596-2023-07271
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013235
UDI-Public00813024013235
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/09/2023
Device Model Number106531US
Device Lot Number7398802
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
Patient Weight89 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-