Manufacturer's investigation conclusion: the reported events of dim pump running leds and the audio being muffled were not confirmed.Additional information provided communicated that no products would be returned for analysis.The root cause of the reported events could not be conclusively determined through this analysis.The device history records were reviewed and the records revealed that the heartmate 3 system controller, serial number (b)(6), was manufactured in accordance with manufacturing and qa specifications.The system controller was shipped to the customer on 09mar2020.Heartmate 3 instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate 3 patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment, including the controller.Heartmate 3 patient handbook section 6 ¿ "caring for the equipment" describes how to care for and clean all equipment, including the system controller.This section instructs the user to, at least monthly, inspect the system controller¿s audio sounders for dirt or grease.If a change in tone or in loudness is noticed during a system controller self-test, the audio speaker sockets may be obstructed.Audio speaker sockets may be cleaned using a small cotton swab that is moistened (not dripping) with rubbing alcohol.Never insert anything sharp (like a toothpick or pin) into the sounder holes.This can damage the speakers inside.The heartmate 3 patient handbook and the heartmate 3 instructions for use caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|