COLOPLAST A/S ALTIS SINGLE INCISION SLING SYSTEM; FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER Back to Search Results
Catalog Number 519650
Device Problem Material Frayed (1262)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to available information, this device required replacement due to shredding.The device was implanted and when the surgeon went to anchor it, the end was found to be shredded.The device was replaced.No other adverse patient effects were reported.
 
Manufacturer Narrative
An altis mesh was received for evaluation.Examination of the mesh revealed the dynamic anchor and suture were detached and not returned.Examination of the mesh where the dynamic suture was attached revealed the welded area of suture.The static anchor and suture were detached and not returned.Blood residue was noted on the mesh.The information received indicated the dynamic anchor and suture detached during the procedure.Quality confirmed the detached dynamic anchor.As the dynamic anchor and suture were not received, it was concluded that the detachment of the dynamic anchor most likely occurred during the tensioning part of the procedure.Detail was not provided as if there were any issues that may have occurred prior to the detachment.A review of the device history record by the contract manufacturer confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Event Description
According to available information, this device required replacement due to shredding.The device was implanted and when the surgeon went to anchor it, the end was found to be shredded.The device was replaced.No other adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTIS SINGLE INCISION SLING SYSTEM
Type of Device
FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17954596
MDR Text Key325911432
Report Number2125050-2023-01254
Device Sequence Number1
Product Code PAH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number519650
Device Lot Number8921905_5196502400
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/18/2023
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-