On sunday september 17th the in-house hospital team had been running experimental clinical beams with the versa hd linac (outside of clinical hours & not treating patients).Having experienced repeated inhibits of low dose rate during the day, at 6:40 p.M the dose rate of the beam dropped from 570 to 430.The hospital team started investigating the low dose fault.Elekta received a call from the hospital engineer requesting assistance in the repair.The versa hd linac was running in service mode with interlocks overridden.The hospital engineers when joined by elekta service engineer at 8:10 pm, they reported no obvious abnormalities when viewing the relevant parameters.The machine was shut down and restarted at about 8:20 p.M.(the versa hd was operating in ctrl t interlock group & service mode with all interlocks overridden).The machine still reported the same fault, low dose rate.The elekta engineer followed by the hospital engineer went to the spare parts room on the 2nd floor to obtain measurement equipment in preparation for performing measurement tests on the versa hd, about 2-3 minutes later back to the operation room, ht psu ot and other faults, were displayed, resetting the equipment did not remove the inhibit.The equipment was shut down and the engineers prepared to enter the computer room to troubleshoot the fault, entering the computer room found that there was obvious smoke.
|