DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-45
Device Problems Failure to Run on Battery (1466); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported during daily maintenance inspection the cs300 intra-aortic balloon pump (iabp) ups function cannot be used, the transfer function cannot be used.There was no patient involvement.
 
Manufacturer Narrative
A getinge field service engineer (fse) reported that the ups battery was replaced and the ups function of the equipment returned to normal.The unit was tested according to the factory requirements and all tests passed.The unit was then returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key17957644
MDR Text Key325946949
Report Number2249723-2023-04501
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108575
UDI-Public10607567108575
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-45
Device Catalogue Number0998-00-3023-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-