It was reported that the patient had an outflow obstruction that required surgical intervention on (b)(6) 2023.A computed tomography scan (ct) had confirmed the obstruction was on the outside of the graft at the bend relief.The patient also showed symptoms of pulmonary and venous congestion, organ failure, and was poorly perfused.After the surgery, the patient still experienced low flows.The patient's status improved, but they remained in the intensive care unit (icu).A log file analysis observed a backup battery fault that occurred on (b)(6) 2023, and low flow alarms occurred on (b)(6) 2023.
|
Manufacturer's investigation conclusion: the reported extrinsic outflow graft obstruction could not be confirmed.A specific cause for the reported obstruction could not be conclusively determined through this evaluation.Additionally, review of the submitted log files confirmed low flow alarms.A specific cause for the reported events as well as a direct correlation between the alarms the outflow graft obstruction could not be conclusively determined through this evaluation.The controller event log files contained events from 06oct2023 through 11oct2023 and 30oct2023.Multiple low flow alarms were captured on (b)(6) 2023.Some of the low flow events captured on 06oct2023 were associated with elevated pi values.The pump appeared to function as intended at the set speed.The patient remains ongoing on heartmate (hm) 3 left ventricular assist system (lvas), serial number mlp-(b)(6) and no further related events have been reported at this time.The relevant sections of the device history records for mlp-(b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate 3 lvas instructions for use (ifu), and the heartmate 3 patient handbook, are currently available.Section 1 of the ifu, ¿introduction¿, addresses all pump parameters, including pump speed, flow, and pulsatility index (pi).In reference to pi, the ifu explains that pi calculation represents cardiac pulsatility and pi values typically range from 1 to 10.In general, the magnitude of the pi value is related to the amount of assistance provided by the pump.Higher values indicate more ventricular filling and higher pulsatility (i.E.The pump is providing less support to the left ventricle).Section 4 of the ifu, ¿system monitor¿, describes the pump flow display and the hazard alarms.Per design, when the estimated flow value is calculated at less than 2.5 lpm, a low flow status is posted to the log file.If the flow remains below 2.5 lpm for 10 seconds, a low flow hazard alarm is triggered.Section 4 also explains that changes in patient condition can result in low flow.Section 5 of the ifu, ¿surgical procedures¿, outlines considerations for pump placement and provides instructions regarding the preparation, installation, and orientation of the sealed outflow graft.Section 5, under "attaching the sealed outflow graft to the pump," instructs the user to verify that the graft is not twisted or kinked by checking the position of the black line on the graft above and below the bend relief and ensuring that the line is straight.Section 5 of the ifu, under "preimplant procedures" and "implant procedures" cautions the user: "the sealed outflow graft must not be kinked or positioned where it could abrade against a pump component or body structure" and "stretch the sealed outflow graft completely prior to measuring and cutting the graft to the appropriate length." section 5 of the patient handbook, ¿alarms and troubleshooting¿, and section 7 of the ifu, ¿alarms and troubleshooting¿, provides information on all system alarm conditions as well as the appropriate actions associated with each condition.A section on ¿handling emergencies¿ is also provided in the patient handbook.No further information was provided.The manufacturer is closing the file on this event.
|