U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION 2 ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION 2 ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX520H11C
Device Problem Temperature Problem (3022)
Patient Problem Burn(s) (1757)
Event Date 05/01/2022
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device not returned to manufacturer.
 
Event Description
The patient reported to philips that while using the dream station 2 advanced auto cpap device and mask was thermal smell and burnt his nose.The device has not yet been returned to the manufacturer for evaluation.Three attempts to have the device returned for evaluation and investigation were unsuccessful.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Manufacturer Narrative
The patient previously reported to philips that while using the dream station 2 advanced auto cpap device and mask was thermal smell and burnt his nose.The previous report included serious injury for the type of reported complaint, which is incorrect.The following sections were corrected: b1, b2, h1 and health impact grid.This report is being filed to correct this information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION 2 ADV AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key17958759
MDR Text Key325965301
Report Number2518422-2023-27126
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200480
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX520H11C
Device Catalogue NumberDSX520H11C
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/24/2023
Initial Date FDA Received10/18/2023
Supplement Dates Manufacturer Received01/24/2023
Supplement Dates FDA Received12/07/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/29/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age83 YR