Model Number CDDRA500Q
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Incorrect Interpretation of Signal (1543); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2023
Event Type  Injury  
Event Description
It was reported during remote follow up that the implantable cardioverter defibrillator exhibited inappropriate shock due to incorrect interpretation of signal.This misinterpretation was caused by a redefinition of active therapy zones that was triggered as a result of under-sensing on the discrimination channel.The device was successfully reprogrammed.The patient was stable and asymptomatic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17959529
MDR Text Key325972610
Report Number2017865-2023-50103
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model NumberCDDRA500Q
Device Lot NumberS000083850
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA; OPTISENSE
Patient Outcome(s) Other;
Patient Age62 YR
Patient SexFemale
-
-