• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 21-7047-24
Device Problem Failure to Deliver (2338)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Manufacturer Narrative
Other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.D4: lot and expiration date unknown; h4 unknown.
 
Event Description
It was reported the disposable was in a device that ran for 168 hours and did not dispense any drug or alarm.No adverse patient effects were reported by the customer.
 
Manufacturer Narrative
Other text: d3, g1,2 email is: regulatory.Responses@icumed.Com.One device was returned for evaluation.Visual inspection revealed no obstructions, damaged, broken, or other defects detected.Functional testing was performed but the reported issue could not be replicated.The root cause could not be determined.No further actions were taken.No lot number was provided; therefore, a history record review could not be conducted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17959918
MDR Text Key325976647
Report Number3012307300-2023-09742
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586020315
UDI-Public10610586020315
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K942046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number21-7047-24
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/02/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2023
Initial Date FDA Received10/18/2023
Supplement Dates Manufacturer Received11/27/2023
Supplement Dates FDA Received12/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-