• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problems Break (1069); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Event Description
It was reported that a device fracture occurred.The stenosed target lesion was located in the severely tortuous and calcified artery.A 10mmx3.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the balloon catheter shaft was broken.There were no patient complications reported post procedure.
 
Manufacturer Narrative
The device was returned for analysis.Visual, tactile, and microscopic analysis were performed on the device.A visual examination of the balloon identified no damages.Multiple hypotube kinks were noted along the shaft of the device, a midshaft kink was 2.5cm from the port exchange and a break was identified, 27.5cm distal to the distal end of the strain relief.No kinks or damages.A detailed microscopic examination of the balloon material identified no tears or pinholes in the balloon.All blades were fully bonded on the balloon and did not exhibit any signs of damage.No issues identified during examination of the extrusion shaft.Tip showed no signs of tip damage.A microscopic examination of the proximal and distal markerbands identified no damage.Multiple hypotube kinks were noted along the shaft of the device, a midshaft kink was 2.5cm from the port exchange and a break was identified, 27.5cm distal to the distal end of the strain relief.
 
Event Description
It was reported that a device fracture occurred.The stenosed target lesion was located in the severely tortuous and calcified artery.A 10mmx3.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the balloon catheter shaft was broken.There were no patient complications reported post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17960276
MDR Text Key326849853
Report Number2124215-2023-54129
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0031151114
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-