• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; NO MATCH Back to Search Results
Model Number CDDRA500Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Event Description
Related manufacture reference number: (b)(4).It was reported that the patient was symptomatic of perforation caused by the right ventricular lead.During reposition, it was noted that the helix of the right ventricular lead failed to extend.The reported lead was explanted and a new lead was used.During implant of the new right ventricular lead, it was noted that the set screw of the implantable cardioverter defibrillator header was stripped and the new lead could not be screwed in.The implantable cardioverter defibrillator was explanted and a the procedure was completed with a new implantable cardioverter defibrillator on (b)(6)2023.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of a set screw anomaly was confirmed.As received, a device was returned for analysis.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.The ventricular set screw hex cavity was found to be stripped.The cause of the stripped set screw hex cavity was procedure related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17962302
MDR Text Key326033594
Report Number2017865-2023-50139
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Lot NumberA000145223
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL LEAD
Patient Age76 YR
Patient SexFemale
-
-