Catalog Number 09043284190 |
Device Problem
Optical Problem (3001)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The analyzer serial number was not provided.The patient sample was requested for investigation.The investigation is ongoing.
|
|
Event Description
|
There was an allegation of questionable elecsys ft4 iv assay results for 1 patient sample on a cobas e 601 analytical unit.The initial ft4 result was 30 pmol/l.The sample was sent to another laboratory and tested on an abbott architect.The abbott ft4 result was 17.2 pmol/l.The elecsys ft4 reference range is 11.4 - 22 pmol/l.The abbott ft4 reference range is 8.5 - 22.5 pmol/l.
|
|
Manufacturer Narrative
|
The investigation did not identify a product problem.It was determined that the result differences generated between the different methods are consistent with methodological differences.
|
|