• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2023
Event Type  malfunction  
Event Description
Philips received a complaint from the customer biomed reporting a machine and proximal pressure sensors failed message displayed on the v60 ventilator.The device was not in clinical use.There was no report of harm.A philips remote service engineer (rse) evaluated the issue with the biomed engineer (bme) and confirmed the reported issue.The diagnostic code for machine and proximal pressure sensors failed was confirmed in the device event log.The bme reported the issue occurred after the air flow sensor assembly was replaced.The rse advised the customer bme to verify the data acquisition (da) to motor controller (mc) cable was properly seated.After the bme reseated the cable, the issue was resolved.The device was operational after repairs were completed.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17967027
MDR Text Key326155547
Report Number2518422-2023-27333
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/21/2023
Initial Date FDA Received10/19/2023
Date Device Manufactured12/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-