The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint, and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history record) for the libre sensor kits were reviewed, and the dhrs showed the libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
A healthcare professional (hcp), on behalf of the customer, reported they experienced a skin reaction while wearing an adc device and experienced irritation and skin allergy at the sensor site.The customer had contact with a healthcare professional and was prescribed unspecified steroid cream for treatment.There was no report of death or permanent impairment associated with this event.
|