• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SUNRISE MEDICAL (US) LLC JAY SEATING CUSHION; WHEELCHAIR CUSHION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SUNRISE MEDICAL (US) LLC JAY SEATING CUSHION; WHEELCHAIR CUSHION Back to Search Results
Model Number JAY UNION
Device Problems Degraded (1153); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Event Description
The dealer reports that the end user informed her that the seated part of the cushion is sinking in and is uncomfortable.The dealer has requested a warranty for this cushion.No user injuries or adverse impact were reported.
 
Manufacturer Narrative
Background information: the age of the wheelchair cushion at the time of the complaint was approximately 1 year and 6 months.The expected lifetime of a wheelchair cushion is 2 years.Jay union cushion owner's manual, rev a, page 2 states: "warning! prior to prolonged sitting, any cushion should be tried for a few hours at a time while a clinician inspects your skin to ensure that red pressure spots are not developing.You should regularly check for skin redness.The clinical indicator for tissue breakdown is skin redness.If your skin develops redness, discontinue the use of the cushion immediately and see your doctor or therapist.The jay union cushion is designed to help reduce pressure.However, no cushion can completely eliminate sitting pressure or prevent pressure sores.The jay union cushion is not a substitute for good skin care including proper diet, cleanliness, and regular pressure relief." jay union cushion owner's manual, rev a, page 3 states: "easy maintenance & cleaning: regular cleaning and maintenance may help extend the life of your cushion.During cleaning, component inspection is recommended.Check the cover for tears and excessive wear and/or any other abnormalities.Check the foam base to ensure foam consistency." discussion: the most probable cause for "bottoming out" for a foam cushion is degradation, or the foam losing its intended shape when being used by the end user.This may cause the cushion to no longer provide sufficient seating and positioning support necessary for the user's specific needs.The dealer reports that the end user informed her that the seated part of the cushion is sinking in and is uncomfortable.As stated in the jay union cushion owner's manual, the cushion should be inspected to check for abnormalities to the foam consistency.If proper support is not being provided, the end user should discontinue using the cushion and immediately contact their sunrise medical supplier.Conclusion: the information in the complaint indicates a reduction of pressure relieving capabilities in the pelvic well area related to deformation of the foam material.The reported event is most likely a product malfunction.The product in question met all product specifications before release for distribution at the time of shipping to the customer.The relationship between the reported event and the device for failing to meet life expectancy is not known.This device is used for treatment, not diagnosis.The failure mode has been previously reported per 21 cfr 803.50.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAY SEATING CUSHION
Type of Device
WHEELCHAIR CUSHION
Manufacturer (Section D)
SUNRISE MEDICAL (US) LLC
2842 n business park avenue
fresno CA 93727 1328
Manufacturer (Section G)
SUNRISE MEDICAL (US) LLC
2842 n business park avenue
fresno CA 93727 1328
Manufacturer Contact
christian stephens
2842 n business park ave
fresno, CA 93727
5592942374
MDR Report Key17970223
MDR Text Key326141219
Report Number9616084-2023-00062
Device Sequence Number1
Product Code IMP
UDI-Device Identifier00016958056402
UDI-Public00016958056402
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberJAY UNION
Device Catalogue NumberJAY
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/19/2023
Initial Date FDA Received10/19/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/18/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexMale
Patient Weight51 KG
-
-