• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 GLIDEPATH; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 GLIDEPATH; DIALYSIS CATHETER Back to Search Results
Catalog Number 6393190
Device Problems Fluid/Blood Leak (1250); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  malfunction  
Event Description
It was reported that three months post dialysis catheter placement, the extension part of the device was allegedly broken.It was further reported that the extension part of the catheter allegedly leaked.Reportedly, the catheter was removed and replaced.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 12/2024) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were provided for review.The photo shows the glidepath catheter and a crack was noted on the extension led of the red luer proximal to the clamp.Therefore, the investigation is confirmed for the reported fracture issue.However, the investigation is inconclusive for the reported fluid leak as no objective evidence was provided for review.The definitive root cause could not be determined based upon the available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 12/2024), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that three months post a dialysis catheter placement, the extension part of the device was allegedly broken.It was further reported that the extension part of the catheter allegedly leaked.Reportedly, the catheter was removed and replaced.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDEPATH
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17974228
MDR Text Key326144793
Report Number3006260740-2023-04782
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741138492
UDI-Public(01)00801741138492
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K051748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6393190
Device Lot NumberREHP3669
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-