Catalog Number 2C8419
Device Problems Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2023
Event Type  malfunction  
Event Description
It was reported that the tubing of a clearlink system solution set had a small hole and leaked.The leak was from a mid-point area between the first y-site and the area that connects to the patient's saline lock.This was discovered during patient infusion with medication containing iron sucrose.To resolve the event, the infusion was stopped and tubing changed.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional underwater pressure test was performed which revealed a leak in the tubing at a location approximately 70cm away from a tubing component.The reported condition was verified.The cause of the condition was determined to be due to a minor rough or sharp edge on the two piece gripper which caused a cut on the tubing during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO-VENT CLEARLINK LUER ACTIVATED VALVE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17977194
MDR Text Key326397308
Report Number1416980-2023-05367
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048956
UDI-Public(01)00085412048956
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112893
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8419
Device Lot NumberSR23D19007
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/26/2023
Initial Date FDA Received10/20/2023
Supplement Dates Manufacturer Received11/15/2023
Supplement Dates FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IRON SUCROSE.
-
-