Catalog Number D139505
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2023
Event Type  malfunction  
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot micro¿ catheter for which biosense webster¿s product analysis lab (pal) identified a cut on pebax with reddish-brown material inside and internal parts exposed.It was initially reported by the customer that the catheter force sensor was not working properly.Catheter force information was inconsistent despite numerous resets.Qmod+ points were all indicated as unstable.Multiple errors were appearing on the carto every time the catheter was ablating.Changing the cable did not solve the issue.Changing the catheter solved the issue.There was no patient consequence.The customer¿s reported force issue is not considered to be mdr reportable since the potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.On (b)(6) 2023, the bwi pal revealed that a visual inspection of the returned device found a cut on pebax with reddish-brown material inside and internal parts exposed.These findings were reviewed and assessed the issue of ¿a cut¿ in the pebax as an mdr reportable malfunction since the integrity of the device has been compromised.
 
Manufacturer Narrative
Device evaluation details: the device was returned to biosense webster (bwi) for evaluation and the evaluation has been completed.Visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed a cut on pebax with reddish-brown material inside and internal parts exposed; however, the cut could be related to the handling since in the process there are control inspection points to avoid this kind of issue; however, this could not be conclusively determined.The magnetic and force feature was tested, and no errors were observed.The force values and the vector were observed within specifications.The blood found inside the pebax area may contribute to the force issue.A manufacturing record evaluation was performed for the finished device, and no internal action was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO¿ CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17978028
MDR Text Key326389426
Report Number2029046-2023-02390
Device Sequence Number1
Product Code OAE
UDI-Device Identifier10846835016758
UDI-Public10846835016758
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD139505
Device Lot Number30962502L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2023
Date Manufacturer Received09/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM; UNKNOWN CABLE
-
-