RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor auto zero failed error code.The device was reported to be in use at the time of the reported problem.No patient harm reported.The customer biomedical engineer (bme) informed the remote service engineer (rse) that they replaced a part on the gas delivery system (gds).The bme stated that they moved the data acquisition (da) printed circuit board assembly (pcba) to another device and the issue followed the da pcba.The rse recommended that the customer perform the following troubleshooting steps: verify the pin alignment between the solenoids and the da pcba, replace the proximal autozero solenoids 3 and 4, and finally replace the da pcba.The rse provided the customer with the part information for the da pcba.This investigation is ongoing.
 
Manufacturer Narrative
Multiple good faith efforts (gfe) were attempted to obtain confirmation of the part order, part replacement, and resolution.No response or further information was received from the customer.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17979739
MDR Text Key326197621
Report Number2518422-2023-27473
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/20/2023
Date Device Manufactured05/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-