• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; OES THORACOSCOPE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; OES THORACOSCOPE SYSTEM Back to Search Results
Model Number WA53000A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: the eyepiece was broken.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the following led to the malfunction: identified fault patterns are most likely attributable to the use of excessive force by the customer (impact, drop, fall).The following non-reportable malfunctions were found during the device evaluation: the light guide bundle was damaged.Olympus will continue to monitor field performance for this device.
 
Event Description
The customer reported to olympus that the telescope, 10 mm, 0°, hd, quick lock, autoclavable had the eyepiece broken.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
OES THORACOSCOPE SYSTEM
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17980433
MDR Text Key326359032
Report Number9610773-2023-03029
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052757
UDI-Public04042761052757
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K912362
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53000A
Device Lot Number638647
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/27/2023
Initial Date FDA Received10/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/19/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-