OLYMPUS WINTER & IBE GMBH TELESCOPE, 10 MM, 0°, HD, QUICK LOCK, AUTOCLAVABLE; OES THORACOSCOPE SYSTEM
|
Back to Search Results |
|
Model Number WA53000A |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation, and the customer¿s allegation was confirmed.The evaluation found the following reportable issues: the eyepiece was broken.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the following led to the malfunction: identified fault patterns are most likely attributable to the use of excessive force by the customer (impact, drop, fall).The following non-reportable malfunctions were found during the device evaluation: the light guide bundle was damaged.Olympus will continue to monitor field performance for this device.
|
|
Event Description
|
The customer reported to olympus that the telescope, 10 mm, 0°, hd, quick lock, autoclavable had the eyepiece broken.There were no reports of patient harm.
|
|