Model Number U228 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439); Connection Problem (2900)
|
Patient Problems
Presyncope (4410); Asystole (4442)
|
Event Date 10/10/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the product has not been returned.If the product is returned, analysis will be performed, and this report will be updated at that time.
|
|
Event Description
|
It was reported that this patient attended the clinic with symptoms of presyncope.Upon device interrogation, it was noted that this cardiac resynchronization therapy pacemaker (crt-p) system had triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) pacing impedance measurements greater than 3000 ohms.Additionally, episodes with oversensing of noise on the ventricular channel were found, which included periods of greater than 14 seconds of asystole.As a result, the patient was admitted to the hospital and a new non-boston scientific rv lead was implanted with evidence of optimal electrical measurements on the pacing system analyzer, but when the physician connected this lead to the already implanted crt-p, no change in the pacing morphology could be observed on the electrocardiogram, and high out of range pacing impedance measurements and noise were present.Further troubleshooting was performed, an after manipulating the device, it was noted that the noise improved but it was being oversensed, and the pacing impedance was still high out of range, measuring 2400 ohms.Consequently, the physician made the decision to explant the crt-p device and a non-boston scientific product was implanted instead as a device header connection problem was suspected.The procedure was completed successfully, and no additional adverse patient effects were reported.The device is expected to be returned for analysis.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.At this time, the product has not been returned.If the product is returned, analysis will be performed, and this report will be updated at that time.
|
|
Event Description
|
It was reported that this patient attended the clinic with symptoms of presyncope.Upon device interrogation, it was noted that this cardiac resynchronization therapy pacemaker (crt-p) system had triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) pacing impedance measurements greater than 3000 ohms.Additionally, episodes with oversensing of noise on the ventricular channel were found, which included periods of greater than 14 seconds of asystole.As a result, the patient was admitted to the hospital and a new non-boston scientific rv lead was implanted with evidence of optimal electrical measurements on the pacing system analyzer, but when the physician connected this lead to the already implanted crt-p, no change in the pacing morphology could be observed on the electrocardiogram, and high out of range pacing impedance measurements and noise were present.Further troubleshooting was performed, an after manipulating the device, it was noted that the noise improved but it was being oversensed, and the pacing impedance was still high out of range, measuring 2400 ohms.Consequently, the physician made the decision to explant the crt-p device and a non-boston scientific product was implanted instead as a device header connection problem was suspected.The procedure was completed successfully, and no additional adverse patient effects were reported.The device is expected to be returned for analysis.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.At this time, the product has not been returned.If the product is returned, analysis will be performed, and this report will be updated at that time.
|
|
Event Description
|
It was reported that this patient attended the clinic with symptoms of presyncope.Upon device interrogation, it was noted that this cardiac resynchronization therapy pacemaker (crt-p) system had triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) pacing impedance measurements greater than 3000 ohms.Additionally, episodes with oversensing of noise on the ventricular channel were found, which included periods of greater than 14 seconds of asystole.As a result, the patient was admitted to the hospital and a new non-boston scientific rv lead was implanted with evidence of optimal electrical measurements on the pacing system analyzer, but when the physician connected this lead to the already implanted crt-p, no change in the pacing morphology could be observed on the electrocardiogram, and high out of range pacing impedance measurements and noise were present.Further troubleshooting was performed, an after manipulating the device, it was noted that the noise improved but it was being oversensed, and the pacing impedance was still high out of range, measuring 2400 ohms.Consequently, the physician made the decision to explant the crt-p device and a non-boston scientific product was implanted instead as a device header connection problem was suspected.The procedure was completed successfully, and no additional adverse patient effects were reported.The device is expected to be returned for analysis.Several attempts were made asking for product return for analysis.Evidence suggests that this device was sent to boston scientific, but further investigation to locate the device has been unsuccessful.If the device is returned in the future, technical analysis will be performed, and this report will be updated at that time.
|
|
Event Description
|
It was reported that this patient attended the clinic with symptoms of presyncope.Upon device interrogation, it was noted that this cardiac resynchronization therapy pacemaker (crt-p) system had triggered a lead safety switch (lss) due to a high out of range right ventricular (rv) pacing impedance measurements greater than 3000 ohms.Additionally, episodes with oversensing of noise on the ventricular channel were found, which included periods of greater than 14 seconds of asystole.As a result, the patient was admitted to the hospital and a new non-boston scientific rv lead was implanted with evidence of optimal electrical measurements on the pacing system analyzer, but when the physician connected this lead to the already implanted crt-p, no change in the pacing morphology could be observed on the electrocardiogram, and high out of range pacing impedance measurements and noise were present.Further troubleshooting was performed, an after manipulating the device, it was noted that the noise improved but it was being oversensed, and the pacing impedance was still high out of range, measuring 2400 ohms.Consequently, the physician made the decision to explant the crt-p device and a non-boston scientific product was implanted instead as a device header connection problem was suspected.The procedure was completed successfully, and no additional adverse patient effects were reported.The device is expected to be returned for analysis.Several attempts were made asking for product return for analysis.Evidence suggests that this device was sent to boston scientific, but further investigation to locate the device has been unsuccessful.If the device is returned in the future, technical analysis will be performed.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.The device was returned for analysis.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Pin gage testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.The associated investigation also determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.At this time, the product has not been returned.If the product is returned, analysis will be performed, and this report will be updated at that time.
|
|