ALERE SAN DIEGO, INC. HENRY SCHEIN HCG URINE CASSETTE; VISUAL, PREGNANCY HCG, PRESCRIPTION USE
|
Back to Search Results |
|
Model Number FHC-102 |
Device Problem
False Positive Result (1227)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/26/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported receiving false positive hcg results while testing at least four patients using the henry schein hcg urine cassettes.The customer provided no patient information.Additionally, no information was provided regarding sample appearance, sample handling and testing technique.No further information was provided.
|
|
Manufacturer Narrative
|
Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded valid negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: - very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.- a number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.- this test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.H3 other text: although requested, device return is not expected.
|
|
Event Description
|
The customer reported receiving false positive hcg results while testing at least four patients using the henry schein hcg urine cassettes.The customer provided no patient information.However, new information received on 23oct2023 noted the appearance of the samples were clear and yellow.The urine samples were allowed to equilibrate to room temperature prior to testing and the tests were out of the foil pouch for a maximum of 3 minutes prior to testing.A timer was not used; however, the customer indicated the results were read at 3 minutes and 2 of the tests yielded faint test lines and the others yielded very bold test lines.No further information was provided.
|
|
Manufacturer Narrative
|
B5 - describe event or problem: the customer provided additional information on 23oct2023.Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded valid negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: - very low levels of hcg (less than 50 miu/ml) are present in urine specimen shortly after implantation.However, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning urine specimen collected 48 hours later.- a number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.- this test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.H3 other text : although requested, device return is not expected.
|
|