• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; IMPLANTABLE LEAD Back to Search Results
Model Number 4543
Device Problem Failure to Capture (1081)
Patient Problems Bradycardia (1751); Low Blood Pressure/ Hypotension (1914)
Event Date 09/29/2023
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this left ventricular (lv) lead showed loss of capture (loc) on the presenting electrogram (egm).The patient also complained of hypotension at home and has seen bradycardia on the presenting.The physician was relayed that the lower rate limit is programmed at 40 bpm, this correlates with heart rate of 47 bpm seen on external telemetry monitor.The patient was seen in the emergency room.The lv output was increased to get consistent capture.Also, the low-rate limit was increased.The pressure immediately improved.The lv lead remains in service at this time.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17982885
MDR Text Key326248299
Report Number2124215-2023-58609
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410994
UDI-Public00802526410994
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2010
Device Model Number4543
Device Catalogue Number4543
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/31/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age87 YR
Patient SexMale
-
-