It was reported that this left ventricular (lv) lead showed loss of capture (loc) on the presenting electrogram (egm).The patient also complained of hypotension at home and has seen bradycardia on the presenting.The physician was relayed that the lower rate limit is programmed at 40 bpm, this correlates with heart rate of 47 bpm seen on external telemetry monitor.The patient was seen in the emergency room.The lv output was increased to get consistent capture.Also, the low-rate limit was increased.The pressure immediately improved.The lv lead remains in service at this time.No additional adverse patient effects were reported.
|