• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-3500
Device Problem Peeled/Delaminated (1454)
Patient Problem Foreign Body In Patient (2687)
Event Date 10/05/2023
Event Type  Injury  
Manufacturer Narrative
Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro vh-3500 part of gray tip came off of tip.Tip fell into patient, and was retrieved by device.New device.No delay.No injuries to patient.
 
Manufacturer Narrative
(b)(4).The device was returned to the factory for evaluation on 10/12/2023.An investigation was conducted on 10/12/2023.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed on the device.Evidence of heavy charred material was observed between the jaws.A microscopic evaluation was conducted.The heater wire and gray silicone insulation of the hot jaw were observed to be intact with no visual defects.The insulation of the cold jaw was observed to be peeled and detached, exposing the metal jaw of the bottom half of the jaw.The detached silicone was not returned for evaluation.Based on the returned condition of the device and evaluation results, the reported failure "peeled; jaw" was confirmed.The lot # 3000336708 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key17985712
MDR Text Key326277999
Report Number2242352-2023-00891
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567701250
UDI-Public00607567701250
Combination Product (y/n)N
PMA/PMN Number
K153194
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-3500
Device Catalogue NumberVH-3500
Device Lot Number3000336708
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2023
Initial Date FDA Received10/23/2023
Supplement Dates Manufacturer Received11/15/2023
Supplement Dates FDA Received11/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK.; UNKNOWN.
Patient Outcome(s) Required Intervention;
-
-