• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problems Tidal Volume Fluctuations (1634); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2023
Event Type  malfunction  
Event Description
Iphilips received a complaint by the customer on the v60 indicating that the device was showing no tidal volume data.It is currently unknown if the unit was in patient use at the time of the reported issue.No patient or user harm was reported.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.The customer was sent a quote for onsite service which is currently pending.This investigation is ongoing.
 
Manufacturer Narrative
Philips received a complaint by the customer on the v60 indicating that the device was showing no tidal volume data.It was reported there was no patient involvement at the time the issue was discovered.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem.A field service engineer (fse) went onsite and confirmed the reported issue.The fse determined that a replacement of the air flow sensor assembly was required.The fse replaced the air flow sensor assembly and confirmed the reported issue was resolved.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key17986516
MDR Text Key326289009
Report Number2518422-2023-27588
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/10/2023
Initial Date FDA Received10/23/2023
Supplement Dates Manufacturer Received11/22/2023
Supplement Dates FDA Received11/29/2023
Date Device Manufactured12/08/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-