During attempted implant, blood was found in the body of the left ventricular (lv) lead.Subsequently, there was difficulty when attempting to remove the stylet from the body of the lv lead.A different lv lead was successfully implanted to resolve the event.The patient was stable and there were no adverse consequences.
|
The reported events of difficulty removing the stylet, lead damage and ¿blood flowed back through the tube¿ were confirmed.As received, a complete lead with stuck stylet was returned in one piece.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual examination found blood/body fluids inside the inner coil that could gained access through the distal tip of the lead which is considered normal.The ptfe coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin with the cap and crimp sleeve were found pulled out of the molded connector stretching the inner coil consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead during the procedure.The cause of the reported events of difficulty removing the stylet and lead damage was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin to be pulled out through the connector assembly.
|