• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24653
Device Problems Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  Injury  
Event Description
It was reported that premature deployment occurred.This 7x40, 130 cm eluvia drug-eluting vascular stent system was selected for use in an endovascular therapy procedure.The 99% stenosed, moderately calcified, moderately tortuous lesion was located in the superficial femoral artery.During the procedure, the lesion was pre-dilated; however, deployment failure occurred, and the stent was stretched during the attempted stent placement.The procedure was completed with another of the same device and there was no patient injury.
 
Event Description
It was reported that premature deployment occurred.This 7x40, 130 cm eluvia drug-eluting vascular stent system was selected for use in an endovascular therapy procedure.The 99% stenosed, moderately calcified, moderately tortuous lesion was located in the superficial femoral artery.During the procedure, the lesion was pre-dilated; however, deployment failure occurred, and the stent was stretched during the attempted stent placement.The procedure was completed with another of the same device and there was no patient injury.It was further reported that the previously reported 7x40, 130 cm eluvia drug-eluting vascular stent system was implanted with 20mm of the stent elongated.An additional eluvia stent was implanted.Post dilation was performed with a balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17990334
MDR Text Key326368065
Report Number2124215-2023-58822
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24653
Device Catalogue Number24653
Device Lot Number0031283726
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/06/2023
Initial Date FDA Received10/23/2023
Supplement Dates Manufacturer Received10/26/2023
Supplement Dates FDA Received11/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ELUVIA STENT
Patient Outcome(s) Required Intervention;
-
-