• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF UNIT "ESG-400"; ELECTROSURGICAL GENERATOR ESG-400 Back to Search Results
Model Number WB91051W
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2023
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to an olympus repair center for evaluation and the customer¿s reported issue was confirmed.The device evaluation found that the e433 error was due to a defective generator board.It was also found that the front panel was cracked.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the device evaluation, the cause of the e433 error was due to a defective generator board.An improved generator board was introduced into production in early july 2020.This change occurred after the concerned product was manufactured.In general, the customer is required to check the function of all devices used prior to a procedure.As stated on the ifu (instructions for use manual) and as a preventative measure, the ifu states a suitable replacement device must be provided during an application.Olympus will continue to monitor the field performance of this device.
 
Event Description
The customer reported that generator has an e433 error code and is constantly restarting.The issue was found in preparation for use.No particular procedure or patient involved.There were no reports of harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF UNIT "ESG-400"
Type of Device
ELECTROSURGICAL GENERATOR ESG-400
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17991017
MDR Text Key326393559
Report Number9610773-2023-03043
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04042761076838
UDI-Public04042761076838
Combination Product (y/n)N
Reporter Country CodeHU
PMA/PMN Number
K203682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWB91051W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/30/2023
Initial Date FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-