BOSTON SCIENTIFIC CORPORATION INOGEN CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G141
Device Problems Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 10/14/2023
Event Type  Injury  
Event Description
It was reported that there were concerns this cardiac resynchronization therapy defibrillator (crt-d) delivered inappropriate therapy.The device delivered three bursts of anti-tachycardia pacing (atp) and six shocks.Technical services (ts) reviewed the device data and concurred with the health care professional (hcp) that the therapy appeared to be a result of an atrial driven rhythm.Device reprogramming options and monitoring alternatives were recommended.This device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17991138
MDR Text Key326348449
Report Number2124215-2023-58868
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/13/2020
Device Model NumberG141
Device Catalogue NumberG141
Device Lot Number129149
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/16/2023
Initial Date FDA Received10/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age77 YR
Patient SexMale
-
-