Model Number 1458Q/86 |
Device Problem
Failure to Advance (2524)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/04/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the patient presented for implant procedure.During the procedure, the left ventricular (lv) lead failed to advance through the guidewire.The lv lead was not used and replaced during the procedure.The patient was stable throughout.
|
|
Manufacturer Narrative
|
The reported event of failure to advance lead in the catheter was not confirmed.As received, a complete lead was returned in one piece without the catheter.The catheter insertion test was performed, and the lead could be inserted inside the catheter and removed easily.Electrical testing found no any indication of conductor fractures or internal shorts.Visual and x-ray examinations of the lead did not find any anomalies.
|
|
Manufacturer Narrative
|
Correction for h2 section in mdr.Supposedly need to check for correction box instead of additional information and device evaluation boxes.
|
|
Manufacturer Narrative
|
The reported event of failure to advance guidewire to the lead was not confirmed.As received, a complete lead was returned in one piece.The guidewire used in the field was not returned with the lead for analysis.The guidewire insertion test was performed, and the guidewire could be inserted inside the lead and removed without difficulties.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray examinations of the lead did not find any anomalies.
|
|
Search Alerts/Recalls
|