• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN JOURNEY II CR KNEE FEM COMP OX; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN JOURNEY II CR KNEE FEM COMP OX; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number UNKNOWN
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Joint Laxity (4526)
Event Date 09/28/2023
Event Type  Injury  
Event Description
It was reported that, after a tka surgery performed on an unspecified date, the patient presented to or with an unstable knee.A revision surgery was performed on (b)(6) 2023, and when she was opened it was realized she had been dislocating posteriorly and had worn the lateral condyle down to the substrate.A journey ll cr oxinium femoral component was explanted and replaced with a sz 6 legion revision femur, along with a 5mm posterior medial augment a 5x5 lateral l-wedge, a 4mm offset coupler, a 16x160 press fit stem and a 15mm gen ll constrained insert, with a jll lock.The current patient health status is unknown.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have definitively contributed to the event.The patient's current condition is unknown, and the patient impact beyond the reported knee instability and revision, could not be determined based on the limited information provided.Therefore, no further clinical/medical assessment can be rendered.A review of the instructions for use documents for knee systems revealed in possible adverse effects that dislocation and subluxation can result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Device specific identifiers were not provided.Therefore, an evaluation of the manufacturing records, complaint history review and prior actions review could not be performed.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN JOURNEY II CR KNEE FEM COMP OX
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17994674
MDR Text Key326364433
Report Number1020279-2023-02040
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2023
Initial Date FDA Received10/24/2023
Supplement Dates Manufacturer Received11/15/2023
Supplement Dates FDA Received11/16/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-