Catalog Number 09043284190 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reagent lot number used in the customer's laboratory was requested but not provided.The reagent lot number used in the investigation laboratory's e 801 is 670634.The expiration date is 30-nov-2023.The serial number of the customer's cobas 8000 core unit is (b)(6).The serial number of the investigation laboratory's cobas e 801 is (b)(6).The investigation is ongoing.
|
|
Event Description
|
The initial reporter received questionable elecsys ft4 iv assay results from one patient sample tested on their cobas 8000 core unit and an investigation laboratory's cobas e 801.The reporter suspected the presence of iga-type macro tsh in the patient sample.The patient sample was rerun on a wako accuraseed and an abbott alinity.It was also sent to an investigation laboratory that uses an e 801 for further analysis.Please refer to the attachment (b)(4) for the table containing the highlighted questionable results.
|
|
Manufacturer Narrative
|
From the information provided, a general reagent issue can be excluded.Assays from different manufacturers can generate different results.This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used.The investigation did not identify a product problem.
|
|