• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE Back to Search Results
Catalog Number 09043284190
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Manufacturer Narrative
The reagent lot number used in the customer's laboratory was requested but not provided.The reagent lot number used in the investigation laboratory's e 801 is 670634.The expiration date is 30-nov-2023.The serial number of the customer's cobas 8000 core unit is (b)(6).The serial number of the investigation laboratory's cobas e 801 is (b)(6).The investigation is ongoing.
 
Event Description
The initial reporter received questionable elecsys ft4 iv assay results from one patient sample tested on their cobas 8000 core unit and an investigation laboratory's cobas e 801.The reporter suspected the presence of iga-type macro tsh in the patient sample.The patient sample was rerun on a wako accuraseed and an abbott alinity.It was also sent to an investigation laboratory that uses an e 801 for further analysis.Please refer to the attachment (b)(4) for the table containing the highlighted questionable results.
 
Manufacturer Narrative
From the information provided, a general reagent issue can be excluded.Assays from different manufacturers can generate different results.This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 IV
Type of Device
FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17995708
MDR Text Key326381068
Report Number1823260-2023-03412
Device Sequence Number1
Product Code CEC
UDI-Device Identifier07613336172780
UDI-Public07613336172780
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K220456
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09043284190
Device Lot NumberASKU, 670634
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/24/2023
Supplement Dates Manufacturer Received10/24/2023
Supplement Dates FDA Received11/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age27 YR
Patient SexMale
-
-