Model Number 3854 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the device was fractured.The target lesion was located in the moderately tortuous and moderately calcified artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, upon introduction of the device, it was noted that the device was fractured.The device was removed intact.The procedure was completed with a different device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.A visual examination of the balloon identified no damages.A hypotube break was identified, 78.5cm distal to the distal end of the strain relief and multiple hypotube kinks were noted along the shaft of the device.The shaft polymer extrusion has no kinks or damages.A detailed microscopic examination of the balloon material identified no tears or pinholes in the balloon.All blades were fully bonded on the balloon and did not exhibit any signs of damage.No issues identified during examination of the extrusion shaft.Tip showed no signs of tip damage.The proximal and distal markerbands identified no damage.
|
|
Event Description
|
It was reported that the device was fractured.The target lesion was located in the moderately tortuous and moderately calcified artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, upon introduction of the device, it was noted that the device was fractured.The device was removed intact.The procedure was completed with a different device.No patient complications were reported.
|
|
Search Alerts/Recalls
|