• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number EMAX2PLUS
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported from japan that the motor device rotated in locked position.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.E1: the reporter¿s phone number was not provided.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.(b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the device had a cord and flex circuit damage, was loose, and ran in locked position.It was further determined that the device made an excessive noise, had vibration, a worn bearing, and leaked liquid.It was further determined that the device failed pretest for visual assessments, loctite assessment, no short circuit between phases and connector body, no short circuit between hall sensors and connector body, safety assessment and noise assessment.Therefore, the reported condition of the motor device rotating in locked position was confirmed.The assignable root cause was determined to be due to component failure from wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key17997503
MDR Text Key326415572
Report Number1045834-2023-01498
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384002055
UDI-Public00845384002055
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received10/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2012
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-