MAUDE Adverse Event Report: ST PAUL CADD ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/26/2023
Event Type  malfunction  
Event Description
It was reported that during connection, the pump alarmed completion approximately 11.5 hours early.The pump was programmed at 144 ml at 1.5 ml/hr.When the pump was disconnected, 126.75 ml infused and 17.3 ml remained, however the bag was emptied.No patient injury was reported.
 
Manufacturer Narrative
Other text: a supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.D4: model number, catalog number, lot number, expiration date, udi, g5, and h4 are unknown.
 
Manufacturer Narrative
Other, other text: d3, g1,2 email is: (b)(4).One device was returned for evaluation.Visual inspection revealed no physical damage.Functional testing was performed but the reported issue could not be replicated or confirmed; no discrepancies were detected.The root cause could not be determined.No further action was taken.No lot number was provided; therefore, a history record review could not be conducted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17999554
MDR Text Key326469114
Report Number3012307300-2023-09852
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Returned to Manufacturer10/10/2023
Was the Report Sent to FDA? No
Date Manufacturer Received11/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-