• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134702
Device Problems Contamination /Decontamination Problem (2895); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Manufacturer Narrative
The product has not returned for analysis, however, pictures were provided by the customer.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and a presence of a gel-like substance in the sterile catheter packaging and catheter was found.Presence of a gel-like substance in the sterile catheter packaging, as well as on the catheter at the tubing connection.This presence was noted just as the catheter was about to be given to the doctor by the ward nurse.The packaging appeared to be intact before it was opened.The catheter was not used.Use of another device to complete the procedure.No surgical delay.No patient consequence.The substance was described as a gelatinous, viscous, thick, transparent substance.
 
Manufacturer Narrative
On 8-nov-2023, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Additionally, the manufactured date and expiration date have been provided.Therefore, fields d4.Expiration date and h4.Device manufacture date have been populated.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Manufacturer Narrative
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and a presence of a gel-like substance in the sterile catheter packaging and catheter was found.Presence of a gel-like substance in the sterile catheter packaging, as well as on the catheter at the tubing connection.This presence was noted just as the catheter was about to be given to the doctor by the ward nurse.The packaging appeared to be intact before it was opened.The catheter was not used.Device evaluation details: the product was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection of the returned device was performed following bwi procedures.The device was inspected, and no physical damage was observed.According to the photo provided by the customer, foreign material like-oil was observed into the sterilization package; however, the package was not returned for analysis.Based on the photo provided by the customer, a manufacturing meeting was performed and the investigation results determined the customer complaint is not confirmed to be related to manufacturing.It is important to state that transparent matter reported by the customer is not used at packaging area and the physical sample was not received.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.The issue reported by the customer was confirmed based on the pictures provided by the customer.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key17999749
MDR Text Key326462883
Report Number2029046-2023-02417
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835009781
UDI-Public10846835009781
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134702
Device Lot Number31119150L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/02/2023
Initial Date FDA Received10/24/2023
Supplement Dates Manufacturer Received11/08/2023
03/08/2024
Supplement Dates FDA Received11/09/2023
03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-