ABBOTT QUADRA ASSURA MP ICD; No Match Back to Search Results
Model Number CD3371-40QC
Device Problem Failure to Convert Rhythm (1540)
Patient Problem Arrhythmia (1721)
Event Date 05/25/2023
Event Type  Injury  
Event Description
It was reported that inadequate antitachycardia pacing (atp) resulting in accelerated ventricular tachycardia (vt) was observed on the device.The vt was converted by high voltage defibrillation therapy administered by the device.The physician alleged the event was not due to a device malfunction and the device performed as expected based upon it's programmed settings.No intervention has been performed at this time to resolve the event.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
No Match
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18004859
MDR Text Key326512091
Report Number2017865-2023-50623
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model NumberCD3371-40QC
Device Lot NumberP000102614
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-