DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during preventive maintenance by getinge field service engineer, the cs300 intra-aortic balloon pump (iabp) unit's battery run test failed.There was no patient involvement.
|
|
Manufacturer Narrative
|
To fix the issue, the fse replaced the batteries.The fse then completed the pm with all functional and safety tests to factory specifications.The unit was returned to the customer.
|
|