Related manufacturer reference number:2017865-2023-50665.Related manufacturer reference number:2017865-2023-50667.Related manufacturer reference number:2017865-2023-50668.It was reported that patient's implantable cardioverter defibrillator (icd), atrial lead, right ventricular lead, left ventricular lead was eroded.Wound dehiscence and hematoma was also noted.The device and all leads were explanted.There were no patient consequences.
|
The device was explanted and returned due to erosion.As received, interrogation of the device revealed it was above elective replacement indicator (eri) when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).No anomalies were found.
|