RESMED LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME) Back to Search Results
Model Number 27003
Device Problems Electrical /Electronic Property Problem (1198); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2023
Event Type  malfunction  
Manufacturer Narrative
Visual inspection of the main circuit board revealed a leaky super capacitor.The main circuit board was replaced to address the reported complaint.The device was serviced and tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to super capacitor electrolyte leak.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
During evaluation by a third party service center, an astral device displayed an error message (sf179) related to the super capacitor.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key18007643
MDR Text Key326555492
Report Number3007573469-2023-00611
Device Sequence Number1
Product Code NOU
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)160426(10)1173917
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation 505
Type of Report Initial
Report Date 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/25/2023
Distributor Facility Aware Date09/26/2023
Device Age88 MO
Date Report to Manufacturer10/25/2023
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/25/2023
Date Device Manufactured05/01/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-