• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems Premature Discharge of Battery (1057); Premature Elective Replacement Indicator (1483); Battery Problem: High Impedance (2947)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/02/2023
Event Type  malfunction  
Event Description
It was reported that the patients generator was recently interrogated and showed that it lost a significant amount of battery life even thought it has only been recently implanted.Internal generator data was received and reviewed.No system diagnostic information was available.It was found that the patient's device is prematurely depleting.It was noted that on 10/02/2023 only 7.543% of the battery was consumed meaning 92.457% of the battery should be remaining, not 21%.On 10/02/2023, the pbcr using history was 92.4567176 and the pbcr using vbat was 21.47288.Per the equation, vbat = 3.518778 + (92.4567176)*0.44567 = 44.72396 versus 21.47288 measured ¿pbcr using vbat¿ on the decoder.Since the measured vbat on the decoder is lower than what the formula gives, the device is below the 95% prediction interval and this is evidence that the battery is depleting faster than expected.No other relevant information has been received to date.
 
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanovas employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any defects¿ or malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Additional information was received noting that upon an updated interrogation of the device, the battery has rebounded to 75-100%.Diagnostics were noted to be within normal limits.The battery rebound event was confirmed to be related to a known event cause by the impedance behaviors in the beginning of the battery's life.Internal investigation identified that in some cases, pulse generators reach the 25% battery indicator earlier than expected, and later rebound to 75-100% without any significant programming changes.This behavior was determined to be due to an increased duration of high battery impedance during generator battery beginning-of-life.There is no evidence of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key18011166
MDR Text Key326613106
Report Number1644487-2023-01537
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number106
Device Lot Number7498
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received11/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexMale
-
-