Catalog Number 565352CB1/B
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  Injury  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow-up report will be submitted once the evaluation is complete.
 
Event Description
The account alleges that during an interventional abdominal aortogram procedure, a pigtail catheter tip detached within the patient the physician had acquired left brachial artery access, and during catheter manipulations, under fluoroscopy, the catheter tip detached.The physician used a vascular snare device to successfully externalize the foreign body with no additional patient consequences to report.
 
Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
.IMPRESS®.BRAIDED.5F CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS MEXICO
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
bryson heaton bsn.rn
1600 merit pkwy
south jordan, UT 84095
8012531600
MDR Report Key18011659
MDR Text Key326618412
Report Number3011642792-2023-00069
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00884450709412
UDI-Public884450709412
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number565352CB1/B
Device Lot NumberI2593047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/17/2023
Initial Date FDA Received10/26/2023
Supplement Dates Manufacturer Received11/27/2023
Supplement Dates FDA Received12/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
INTERVENTIONAL GUIDEWIRE
Patient Outcome(s) Required Intervention;
-
-