Model Number ZEPHYR 4.0 EBV
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pneumothorax (2012)
Event Type  Death  
Event Description
The patient had a bronchoscopic lung volume reduction procedure with zephyr valves placed in the right upper lobe (rul) on (b)(6) 2023.It was reported that the patient had a pneumothorax and died after the procedure.A request for a complete narrative and additional details have been requested but not received.The additional information will be submitted in a follow-up report.
 
Event Description
Patient had a history of aortic tavi and stable, homogenous emphysema.Patient was a candidate for bronchoscopic lung volume reduction (blvr) procedure and was treated with valves placed in the right upper lobe (rul) on (b)(6) 2023.The patient experienced a pneumothorax post-procedure, across fissure from the right lower lobe (rll).Patient had difficulty waking up post-procedure and desaturated.A small amount of blood was observed in chest drain.That same evening, patient had a second bronchoscopy procedure and one valve was removed.The pneumothorax did not appear large, but "frothy" blood was present.Patient was sent to the icu.The following day patient was taken to the operating room and a thoracotomy performed and the rll was removed surgically.Patient was kept unconscious as very anxious when awake.The family withdrew resuscitation permission.The patient died on (b)(6) 2023, 10 days after procedure.No other details were made available despite multiple requests for a full narrative.
 
Manufacturer Narrative
Product model information has been made available, but lot numbers/udi were not provided.Pneumothorax is the most common side effect associated with the zephyr valve treatment (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).Targeted lobar deflation likely causes inflation of the ipsilateral lobe, which can result in a tear of the already compromised parenchymal tissue of the emphysematous ipsilateral lobe, resulting in a pneumothorax (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).In the liberate study (ide clinical study used to support pma p180002's approval), 26.6% of the zephyr valve subjects experienced a pneumothorax in the treatment period ([less than or equal to 45 days).These were managed using standard of care procedures as per previously published guidelines (valipour, arschang, et al.Respiration 87.6 (2014): 513-521).In 17.4% of the events, the pneumothorax resolved without any additional intervention with subjects under careful observation.In over half the events (56.5%), the pneumothorax was managed with a chest-tube only.An additional 13% of the events were managed with a chest-tube and the temporary removal of one or more valves, while another 13% of the events were managed with a chest-tube and removal of all the implanted valves.Upon successful resolution of the pneumothorax, removed valves can be replaced.Patients that experienced a pneumothorax experienced clinical benefits of the zephyr valve treatment that were similar to the benefits experienced by patients who did not have a pneumothorax.The zephyr ebv system ifu and pulmonx training program both specifically reference pneumothorax as a known side effect of this procedure and the published guidelines (valipour, arschang, et al."expert statement: pneumothorax associated with endoscopic valve therapy for emphysema-potential mechanisms, treatment algorithm, and case examples." respiration 87.6 (2014): 513-521).The reported event aligns with the experience observed in the liberate clinical study and is an expected side effect to the zephyr valve treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR ENDOBRONCHIAL VALVE
Type of Device
ENDOBRONCHIAL VALVE
Manufacturer (Section D)
PULMONX CORPORATION
700 chesapeake drive
redwood city CA 94063
Manufacturer Contact
mahtab fatemi
700 chesapeake drive
redwood city, CA 94063
6504072823
MDR Report Key18012707
MDR Text Key326628401
Report Number3007797756-2023-00196
Device Sequence Number1
Product Code NJK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P180002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberZEPHYR 4.0 EBV
Device Catalogue NumberEBV-TS-4.0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/29/2023
Initial Date FDA Received10/26/2023
Supplement Dates Manufacturer Received09/29/2023
Supplement Dates FDA Received02/19/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Death;
Patient Age67 YR
-
-