• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DW ARTHROSCOPY FLUID MANAGEMENT DEV; ARTHROSCOPE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DW ARTHROSCOPY FLUID MANAGEMENT DEV; ARTHROSCOPE AND ACCESSORIES Back to Search Results
Model Number DW ARTHROSCOPY FLUID MANAGEMENT DEV
Device Problem Overheating of Device (1437)
Patient Problem Insufficient Information (4580)
Event Date 07/29/2022
Event Type  malfunction  
Event Description
On (b)(6) 2022, it was reported by a sales representative via email that an ar-8330h shaver handpiece caught on fire when it was plugged in and laying on the mayo stand.Shortly after, the room experienced a power outage.An ar-6480 dualwave arthroscopy fluid management system, an ar-8305 synergy resection shaver console, and an ar-9800 synergy rf console were all connected and running during this incident.This was discovered during a procedure.Additional information 10/5/2022: per the sales representative: the hospital¿s biomedical engineer team did their own internal investigation and found that the shaver disposable was bent during the procedure which caused the inner and outer sleeves to create friction and heat and catch fire.On (b)(6) 2023, the complaint reporter provided additional information confirming that the event occurred after the device had been used in the procedure.In addition, it was informed that the disposable was on the mayo stand when someone accidentally stepped on the footswitch and activated the shaver.This information confirmed arthrex's hypothesis that the device had been activated without irrigation, causing the excessive heating of the disposable.On the same day, arthrex assessed the effects of activating disposable attachments without irrigation, affirming that irrigation is essential in maintaining the device at appropriate temperatures during operation.The labeling for disposable attachments, dfu-0215-7, includes two precautions related to running the device without irrigation flow, indicating that doing so will result in irreversible damage, excessive heating, and may cause thermal burns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DW ARTHROSCOPY FLUID MANAGEMENT DEV
Type of Device
ARTHROSCOPE AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18013399
MDR Text Key326732575
Report Number1220246-2023-08397
Device Sequence Number1
Product Code HRX
UDI-Device Identifier00888867039377
UDI-Public00888867039377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDW ARTHROSCOPY FLUID MANAGEMENT DEV
Device Catalogue NumberAR-6480
Device Lot Number21801067
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/29/2022
Initial Date FDA Received10/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-