• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RADICAL-7 HANDHELD; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RADICAL-7 HANDHELD; OXIMETER Back to Search Results
Model Number 25483
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2023
Event Type  malfunction  
Manufacturer Narrative
Additional manufacuring narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event Description
The customer reported the device turns off on its own.There were no patient impact or consequences reported.
 
Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned device was evaluated.The device internal battery was swollen, which prevented it from being charged.The device was not able to be powered on and would not stay powered on if not connected to an electrical power source.
 
Event Description
The customer reported the device turns off on its own.There were no patient impact or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL-7 HANDHELD
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
tahereh sedighi
15750 alton pkwy
irvine, CA 92618
9492977862
MDR Report Key18017008
MDR Text Key327265458
Report Number3019388613-2023-00262
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997008358
UDI-Public00843997008358
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25483
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/03/2023
Initial Date FDA Received10/26/2023
Supplement Dates Manufacturer Received11/27/2023
Supplement Dates FDA Received12/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-