ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problems Laceration(s) (1946); Localized Skin Lesion (4542)
Event Date 09/27/2023
Event Type  Injury  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: canada.Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that during surgery the dermatome created skip lesions.Further strips of skin were obtained causing a larger donor site and increased surgical time (25-30 minutes with patient under anesthesia) as the skin grafts had to be patched together.In such case, an additional skin graft was required from the patient.Due diligence is complete.No additional information is available.No additional consequences have been reported as a result of this malfunction.
 
Event Description
No additional event information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, g3, g6, h2, h3, h6, h10.Review of the most recent repair record could not be completed because the device was not returned for evaluation and repair.Lot/serial identification is necessary for review of device history records, and lot/serial identification was not provided.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18021234
MDR Text Key326725577
Report Number0001526350-2023-01409
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexMale
-
-