Model Number U128
Device Problems High impedance (1291); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2023
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited high out of range pacing impedance that triggered a lead safety switch (lss) on the right atrial (ra) channel.Additionally, there were minute ventilation (mv) artifacts that triggered a signal artifact monitor (sam) episode on the same channel.A device replacement is planned for this patient.The device remains in use.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv) sensor that is related to intermittent increases in impedance measurements.Additionally, engineering analysis and testing of returned products has identified that any repeated, small movements of the lead terminal ring can create wearing of the lead terminal ring and generate microscopic particles, which may accumulate and oxidize over time.This can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurements.Please refer to the description field for more information regarding the specific circumstances of this event.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) exhibited high out of range pacing impedance that triggered a lead safety switch (lss) on the right atrial (ra) channel.Additionally, there were minute ventilation (mv) artifacts that triggered a signal artifact monitor (sam) episode on the same channel.A device replacement is planned for this patient.The device remains in use.No adverse patient effects were reported.Additional information indicates that this device was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE X4 CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18021369
MDR Text Key326733547
Report Number2124215-2023-59940
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/03/2019
Device Model NumberU128
Device Catalogue NumberU128
Device Lot Number720979
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-